Humidity chamber box
Chapter 1: The main technical parameters for SMC-80L-CD
01. |
Equipment performance: Products performed as air-cooled and water-cooled. The temperature is 25°C when machine is no-load. |
02. |
Product Name: Programmable constant temperature and humidity test chamber |
03. |
Machine Model: SMC-80L-CD |
04. |
The dimensions 400 x 400 x500mm (W x D x H) |
05. |
Volume 80L |
06. |
External dimensions of approximately 700 x 1220 x 1450 mm (W x D x H) |
07. |
Temperature range -70 ℃ ~ 180 ℃ |
08. |
Humidity range from 10% to 98% |
09. |
Heating rate from -70 ℃ to 100 ℃ / have been completed within 40 minutes; Load |
10. |
Cooling rate from 25 ℃ to -70 ℃ / have been completed within about 50 minutes; Load |
11. |
Temperature distribution uniformity of ± 1 ℃ |
12. |
Temperature control volatility ± 0.1 ℃ |
13. |
Temperature deviation ± 1 ℃ |
14. |
Temperature stabilization time ≤ 10min; |
15. |
Humidity distribution uniformity ± 2% |
16. |
Humidity control volatility ± 2% |
17. |
Humidity deviation ± 2% |
18. |
Humidity stabilization time ≤ 10min |
19. |
Total weight 400KG for each unit/set |
Chapter 2: Mechanical structure design:
01. |
Inner box material: Heatproof and cold resistance stainless steel plate with circumferential welding (SUS # 304) |
02. |
Outside box material: Using galvanized sheet with advanced powder paint or Heatproof and cold resistant steel plate (SUS # 304) |
03. |
Cold/heat storage material: High efficiency aluminum to store large of heat and cold so that achieve cold and hot exchange. The main structure design: Composed of four parts: Metal plate, Circuit , Frozen water flow . |
04. |
Observation window: Install multi-layer hollow rectangular insulated glass windows (310 × 420mm)and there is energy-efficient LED lighting on the window. |
05. |
Sample bearing course: Sample on one layer is not more than 30 kg |
06. |
Way to move the sample rack: Sample holder is moved up and down with reciprocating linear |
07. |
Insulation: Thermal insulation layer composed of flame fire PU + glass wool insulation (insulation thickness 10.0 cm) |
08. |
Box tray: adjustable shelves and two stainless steel strips (5.0 cm interval between two layout) |
09. |
Machine Positioning: There are installation of four high-loading corner bearing pulleys and PU wheels at the bottom of the machine to facilitate moving the machine. |
10. |
Cabinet Structure: Taiwan imported plane embedded rotatable knob; |
Chapter 3: Temperature and humidity electric circulatory system
01. |
Electric circulatory system: Designed as thermal ,moisture-proof and stainless steel loop extension axis motor |
02. |
Circulating fan: High / low temperature resistance aluminum Sirocco fan |
03. |
Electric storage control: Balanced thermostat with P.I.D and control with S.S.R |
04. |
Microcomputer control structure: Microcomputer control, pre-cooling preheating temperature conversion control, computer calculations can reach high precision and high output power saving . |
05. |
Automatic control: 1)U-shaped fin high-speed heating electric heater; 2)Refrigeration system is completely independent so does not affect the control lines; 3)Calculated with the heat content of the heating rate of test substance. |
06. |
Humidity system: 1) using micro-electronic and humidification systems.2) Humidification barrel made of stainless steel along with water level observation window. 3) Attached overheating and overflow device. 4) Humidification and dehumidification system is completely independent. 5) Water level control with mechanical float raft, to eliminate electronic malfunction.6) Water supply system with automatic replenishment attached filter so that the buyers don’t need to add water manually.7) using expendable recycled water system. |
07. |
Dehumidification system: 1) Independent compression and refrigeration system . 2) Dew point temperature of the evaporator tube containers (ADP) layer contact with dehumidification. |
THE MAIN PARTS FOR SMC-10-CD (1000 L), LIST AS BELOW MENTIONE
Parts Name | Place/Brand | Model | Qty | Original | |
1 | Temperature humidity controller TEMI/KOREA | TEMI-2700 | 1 | KOREA | |
2 | Relays | IDEC | RH1B-UD24 | 6 | JAPAN |
3 | Electromagnetic contactor | SCHNEIDER | LC1D25 | 5 | GERMANY |
4 | Power switch | MINGWEI | NES-100-24 | 1 | TAIWAN |
5 | Expansion valve | DANFOSS | TES2/068Z3403 3 | DENMARK | |
6 | Overtemperature protection | RAINBOW | TS-320S | 2 | KOREA |
7 | Solid State Relays | SIEMENS | 3RF24 | 2 | GERMANY |
8 | Temperature humidity sensors | CAHO | PT100 | 3 | TAIWAN |
9 | Manual valve | DANFOSS | KVR+NRD | 2 | DENMARK |
10 | Chambers material | KONGLI | SUS 304 | 1 | CHINA |
11 | Pressure protection switch | DANFOSS | KP15 | 2 | DENMARK |
12 | Air-break switch | SCHNEIDER | C-65N | 2 | GERMANY |
13 | Switching Power Supply | MITSUBISHI | NF125-CW | 2 | JAPAN |
14 | Cycle motor | TECO | DY-375W | 2 | TAIWAN |
15 | Compressor | TECUMSSEH | TAG-2516Z | 2 | FRANCE |
16 | Condenser | ZHONGLI | ZL-6PH | 1 | TAIWAN |
17 | Evaporator | ZHONGLI | ZL-6PH | 1 | TAIWAN |
18 | Drying screening program | DANFOSS | DML-084S | 2 | MEXICO |
19 | Solenoid valve | DANFOSS | EVR3 | 4 | DENMARK |
20 | Refrigerant | USA Dupont | R404A/R23A | 2 | USA |
21 | Frozen oil | UK Uniqema | RL32H | 1 | UK |
22 | An oil separator | DANFOSS | OUB040B0010 | 2 | DENMARK |
23 | Cooling fan | EBM | EBM-120 | 3 | GERMANY |
24 | Safe door- lock | Shangkun | SK1-200 | 2 | USA |